| Model Number |
42-9431TRAK2 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.Root cause could not be determined.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of (b)(6) 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during insertion of the nasogastric (ng) tube, "resistance was felt at around cardia and it was difficult to insert the tube.The user tried to move the tube in and out a few times bending the tube.The tube was torn and stylet came out from the tear on the tube while the tube was placed in the patient.No patient injury reported." per additional information received on 13 oct 2020, the tube was not flushed with water when removing the stylet.No further information has been provided at this time.
|
| |
|
Search Alerts/Recalls
|