• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR120610
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a lot history review was performed.The sample was returned to the manufacturer for inspection/evaluation.The investigation is confirmed for detachment and break issue.A definitive root cause for the reported event could not be determined.The device is labeled for single use.
 
Event Description
This report summarizes one malfunction.A review of the reported information indicated that model dr120610 pta balloon dilatation catheter allegedly experienced detachment of device or device component and break.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The age, weight and gender of the patient were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens TX 75751
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key10752862
MDR Text Key213648457
Report Number2020394-2020-06241
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741058844
UDI-Public(01)00801741058844
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDR120610
Device Catalogue NumberDR120610
Device Lot Number93QC0085
Initial Date Manufacturer Received 09/30/2020
Initial Date FDA Received10/28/2020
Type of Device Usage N
Patient Sequence Number1
-
-