ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD LEFT SIZE 6; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Swelling (2091); Limited Mobility Of The Implanted Joint (2671)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2020-00281, 0001822565-2020-03649.Dob: (b)(6).Concomitant medical products: tibia cemented 5 degree stemmed left size c, item# 42532006401, lot# 63783320.Articular surface fixed bearing cruciate retaining (cr) left 10 mm height, item# 42511000410, lot# 63596762.Report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation because it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient experienced pain, swelling, stiffness and decreased range of motion which required medication approximately two years post-implantation.There is no additional information at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
No product was returned or pictures provided; therefore, visual and dimensional evaluations could not be performed.Polar study report was provided and reviewed by a health care professional.Review of the available records identified the following: patient experienced severe pain in the medial joint line, lateral joint line and infrapatellar region for more than two years and indicated that the patient was administered pain killers.Stiffness with rom of 85-90 was noted during this two years and occasional swelling was noted too.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.Per package insert 87-6204-022-23 rev a: pain, swelling and poor range of motion is known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED CRUCIATE RETAINING (CR) STANDARD LEFT SIZE 6
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key10752899
Report Number3007963827-2020-00280
Device Sequence Number17015715
Product Code JWH
UDI-Device Identifier00889024230330
UDI-Public(01)00889024230330
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502606001
Device Lot Number63605407
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/19/2020
Supplement Date Received by Manufacturer01/20/2021
Initial Report FDA Received Date10/28/2020
Supplement Report FDA Received Date01/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight98
-
-