| Model Number |
BD-P80040ER |
| Medical Device Problem Codes |
Material Rupture (1546); Material Separation (1562); Use of Device Problem (1670); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
From the investigation of the actual product, we confirmed a loss of about 42 mm in total for the rupture of the balloon part and the balloon part and the tip part.On the other hand, we have confirmed that there is no problem with the device history record.Our balloon part is designed to rupture in the vertical direction regarding pressure from the inside, but it may rupture in the lateral direction when expanding in calcified lesions or tortuous lesions etc.The actual device was broken by pulling it in a state when the balloon portion was ruptured in the lateral direction.This was considered a malfunction caused by the procedure.
|
| |
|
Event or Problem Description
|
|
During a fistulogram angioplasty, you could see the balloon not fully expand even at burst pressure.The balloon burst.Upon removal it was noted that the balloon burst circumferentially vs longitudinally and part of the balloon (nylon) was left in the aneurysmal section of the fistula.The balloon catheter was removed and the patient access site closed.
|
| |
|
Search Alerts/Recalls
|