TITAN MANUFACTURING MEQUON ENDOSKELETON TL INTERBODY SYSTEM; DRIVER, PROSTHESIS Back to Search Results
Model Number 7200-1088
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with degenerative disc disease (ddd) for spinal olif fusion therapy.It was reported that while implanting theimplant, the prong on the all-in-one integrated inserter broke.Surgeon had to remove the broken instrument, take out the broken piece, remove the cage, and implant via a different method.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis #704005827:7200-1088 lot# ee190701 visual and functional testing identified that the tip of the inner shaft has be broken off and was not returned for analysis.The instrument cannot function as intended.This is consistent with overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSKELETON TL INTERBODY SYSTEM
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
TITAN MANUFACTURING MEQUON
6140 w executive dr ste a
mequon WI 53223
MDR Report Key10753326
Report Number3006340236-2020-00009
Device Sequence Number4142095
Product Code HWR
UDI-Device Identifier00191375045908
UDI-Public00191375045908
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K191581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7200-1088
Device Catalogue Number7200-1088
Device Lot NumberEE190701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2020
Supplement Date Received by Manufacturer11/06/2020
Initial Report FDA Received Date10/29/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
-
-