MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Decreased Sensitivity (2534)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during implant the implantable cardiac monitor (icm) experienced sensing issues.It was further reported the icm detected very low r-waves despite repositioning and vectors.The icm was replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key10754489
MDR Text Key213636160
Report Number9614453-2020-03462
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169845749
UDI-Public00643169845749
Combination Product (y/n)N
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2021
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2020
Date Manufacturer Received12/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age75 YR
-
-