| Model Number |
MI1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
Impedances on the concerned right side have changed over time.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode, as consistent with minute device mobility, was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
Impedances on the concerned right side have changed over time.No accident or trauma has been reported.Re-implantation took place on (b)(6) 2020.
|
| |
|