| Model Number |
450952-NL |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the device was in use with patient and leakage was noted.No adverse effects to patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: investigation completed on a smiths breathing|portex general anesthesia circuits.Pictures were received and observe that that a sample p/n 450952-nl, l/n 3918232, the sample was received in used conditions without their original packaging.Sample had pinhole.During functional testing the device did not pass the circuit test leak.Complaint confirmed.Capa 000415 was opened on 22/dec/2017, in order to determine the root cause and implement proper corrective actions for leaking.By the time this report has been completed 10/nov/2020 capa 000415 status is in implementation phase.Corrective actions will be addressed on capa 000415.
|
| |
|
Event or Problem Description
|
|
Investigation completed and summarized in h 10.
|
| |
|