(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
|
(b)(4).Date sent to the fda: 11/10/2020.Additional information: a manufacturing record evaluation was performed for the finished device batch pmm784, and no non-conformances were identified.Additional h3 investigation summary: eight unopened samples of product code vcp719, lot # pmm784 were received for evaluation.During the visual inspection of eight unopened samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The device performed without any defect noted and the actual complaint sample was not returned for analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|