C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE Back to Search Results
Model Number 0035280
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the foreign material (dust adhesion) was found on the syringe.Per follow up on (b)(6) 2020, the foreign material was present inside the syringe upon opening of the package.
 
Additional Manufacturer Narrative
The reported event was confirmed, however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device did not meet the specifications, and was influenced by the reported failure.The product used for the treatment.Visual evaluation of the returned sample noted one opened (within original package) unused irrigation bulb syringe.It was noted that there was a 0.2125" long cardboard-like foreign material taped onto the syringe bulb.A potential root cause for this failure mode could be due to ¿no follow up to the production areas cleaning procedure." the device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labeling review was not performed due to the labeling could not have prevented the reported failure.Correction: d.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the foreign material (dust adhesion) was found on the syringe.Per follow-up via ibc on 13oct2020, the foreign material was present inside the syringe upon opening the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Common Device Name
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10755576
Report Number1018233-2020-20549
Device Sequence Number11028138
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number0035280
Device Catalogue Number0035280
Device Lot NumberNGEN4041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2020
Initial Date Received by Manufacturer 10/13/2020
Supplement Date Received by Manufacturer02/05/2021
Initial Report FDA Received Date10/29/2020
Supplement Report FDA Received Date03/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-