| Model Number |
24700 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event was unknown.Used first date of the month the event was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the wire became stuck inside the balloon.A 200cm v-18 control wire and a 2.5mm x 60mm x 90cm sterling were selected for a procedure.While advancing the balloon on the guidewire, the guidewire became stuck inside the balloon.The device was removed.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned device consisted of a sterling balloon catheter with a guide wire inside the device.There was fluid in the shaft and balloon.The balloon was loosely folded.110 cm of the guide wire was protruding out of the catheter hub, and the guide wire was removed from the sterling catheter during analysis.Analysis of the tip, balloon, inner/outer shaft, and hub included visual and microscopic inspection.Inspection revealed the inner shaft was stretched and split open starting 55mm from the tip of the device.No other damage or defect was found with the returned device.The wire used with the device was used for device-to-device testing.The proximal end of the guide wire was inserted into the tip of the sterling catheter and was able to advance 60mm, then stopped inside the inner shaft damage and could not advance any further.The inner diameter (id) of the device was measured and met the product specification.The guide wire was measured with a calibrated micrometer and was less than.018" meeting specification.
|
| |
|
Event or Problem Description
|
|
It was reported that the wire became stuck inside the balloon.A 200cm v-18 control wire and a 2.5mm x 60mm x 90cm sterling were selected for a procedure.While advancing the balloon on the guidewire, the guidewire became stuck inside the balloon.The device was removed.No patient complications were reported.It was further reported that this was a recanalization procedure for an occlusion in the leg.A 2.5mm x 120mm x 90cm sterling was used to complete the procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that the wire became stuck inside the balloon.A 200cm v-18 control wire and a 2.5mm x 60mm x 90cm sterling were selected for a procedure.While advancing the balloon on the guidewire, the guidewire became stuck inside the balloon.The device was removed.No patient complications were reported.It was further reported that this was a recanalization procedure for an occlusion in the leg.A 2.5mm x 120mm x 90cm sterling was used to complete the procedure.
|
| |
|