SMITH & NEPHEW, INC. DYONICS POWERMINI WITH HAND CONTROLS; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72201500 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
|
|
Event Description
|
It was reported that during set up procedure, the motor was not running.No delay and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of motor stall could not be reproduced.Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions.Unit passed functional tests on the dii and dii eip test control units with footswitch.Motor stall did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|