Model Number MP-KIT-P-BI-00590
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2020
Type of Reportable Event Malfunction
Event or Problem Description
Physician informed fhc that there was an issue with a starfix case.Physician reports that a platform height of 130 was chosen for the plan for this bilateral platform for patient.In the case physician observed a value of t=30 on the platform embossing.Being new to starfix, she followed this label and implanted the lead at 30 without questioning it.In post-operative scans, it was observed that the lead is 10 mm short of her target.The patient impact is that he/she will require a revision which bick plans to do using rosa.Physician informed fhc that they decided to bring the patient back in for a revision on10/1 with the same platform.They re-processed the platform and reattached without issue.Patient did very well during the revision surgery and in post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FHC WAYPOINT STEREOTACTIC SYSTEM
Common Device Name
STEREOTACTIC INSTRUMENT
Manufacturer (Section D)
FHC, INC.
1201 main street
bowdoin ME 04287
MDR Report Key10756025
Report Number3002250546-2020-00006
Device Sequence Number19704581
Product Code HAW
UDI-Device Identifier00873263005903
UDI-Public00873263005903
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K092192
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMP-KIT-P-BI-00590
Device Catalogue NumberMP-KIT-P-BI
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2020
Initial Report FDA Received Date10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-