• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-425-25
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that while deploying the pipeline stent, the middle of the device would not open.Resheathing was done twice, but the issue did not resolve.It was noted the stent was positioned in a bend, more than 50% of the pipeline had been deployed, and no additional steps were taken in an attempt to open the stent.The catheter and pipeline were removed together and discarded.Replacement devices were used to complete the procedure.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Post-procedure angiography showed good results.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm of the ophthalmic artery with a max diameter of 19 mm and a 8 mm neck diameter.It was noted the patient's vessel tortuosity was moderate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10757115
Report Number2029214-2020-01072
Device Sequence Number11733450
Product Code OUT
UDI-Device Identifier00847536016231
UDI-Public00847536016231
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2021
Device Model NumberPED-425-25
Device Catalogue NumberPED-425-25
Device Lot NumberA672266
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2020
Initial Report FDA Received Date10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-