• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6534
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that shaft break occurred.The 95% stenosed target lesion was located in the mildly tortuous and severely calcified mid left anterior descending artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.During removal, the distal portion remained in patient due to a bend.The device was completely removed by trapping the wire and inflating another balloon.The procedure was completed with another of the same device and no further complications were reported.
 
Event or Problem Description
It was reported that shaft break occurred.The 95% stenosed target lesion was located in the mildly tortuous and severely calcified mid left anterior descending artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.During removal, the distal portion remained in patient due to a bend.The device was completely removed by trapping the wire and inflating another balloon.The procedure was completed with another of the same device and no further complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 56.8cm distal of the strain relief.The was blood in the guidewire lumen.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10757988
Report Number2134265-2020-14604
Device Sequence Number11733445
Product Code LOX
UDI-Device Identifier08714729783572
UDI-Public08714729783572
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model Number6534
Device Catalogue Number6534
Device Lot Number0025653715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2020
Initial Date Received by Manufacturer 10/09/2020
Supplement Date Received by Manufacturer11/12/2020
Initial Report FDA Received Date10/29/2020
Supplement Report FDA Received Date12/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-