• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Defective Alarm (1014); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 29oct2020.
 
Event Description
The customer reported that the unit failed with backup alarm test failure.The customer contacted product support and requested for service repair.The customer reported that the unit was not in use on a patient.
 
Manufacturer Narrative
G4:22oct2020.B4:15dec2020.The bench service engineer could not replicate the fault.The central processing unit (cpu) board was replaced as a precaution.Set the sn, options, and total power-on hours.The service engineer performed full performance assurance (pa) testing and extended self-test.All tests passed successfully.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4:06feb2021; b4:15feb2021.Failure analysis on the returned cpu board it was determined that it was ls1 failure.Ls1's built without a date code could be subject to cold joints within ls1 that can result in ls1 failing to sound.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10758704
MDR Text Key215273567
Report Number2031642-2020-03962
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-