Model Number 430463
Medical Device Problem Codes Positioning Failure (1158); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
The pulsar-18 t3 stent system was chosen for the treatment.The stent could not be released and after removal a kink was detected.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the outer shaft has not been retracted and that the stent has not been released.The retractable shaft has separated from the bushing which is connected to the pull wire being intended to release the stent.This most likely occurred during intervention when it was attempted to release the stent.Microscopic inspection revealed a rather small amount of remaining glue on the outside of the retractable shaft and no remaining glue inside the bushing.This finding indicates that the root cause for the complaint event is most likely related to the manufacturing process of a subassembly manufactured by a supplier.However, a strong kink was observed in the device shaft which may have contributed to the complaint event.To prevent recurrence the respective internal departments were informed about this case and we are reviewing our quality systems processes.The supplier of the affected sub assembly was informed about the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSAR-18 T3 6/120/90
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10758855
Report Number1028232-2020-04702
Device Sequence Number19117372
Product Code NIP
UDI-Device Identifier07640130446618
UDI-Public07640130446618
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P160025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/09/2023
Device Model Number430463
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03206006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Received by Manufacturer 10/29/2020
Supplement Date Received by Manufacturer12/04/2020
Initial Report FDA Received Date10/29/2020
Supplement Report FDA Received Date12/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-