| Model Number |
DBP-150SOLID145 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported oad was received at csi for analysis, with the guide wire engaged.Visual examination revealed the guide wire was seized within the driveshaft tip bushing area.Embedded biological material was observed within the tip bushing area.Examination of the area of adhered material did not reveal any damage that would have contributed to the material accumulation.The returned guidewire could not be removed from the driveshaft.When tested, the oad functioned as intended.At the conclusion of the device investigation analysis, the report that the device became stuck on the wire was confirmed.The morphology and exact root cause of the accumulation was unknown.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
|
| |
|
Event or Problem Description
|
|
The diamondback peripheral orbital atherectomy device was operated in the popliteal artery at low speed for three treatment passes.The brake was unlocked and the oad was advanced to the next lesion.Three treatment passes were performed at low speed.As the oad was removed from the vessel, it was stuck on the guide wire and the guide wire moved with the oad.Despite troubleshooting attempts, the oad remained stuck on the guide wire.A second wire was inserted into the guide catheter to maintain wire access and the oad and wire were removed together.The procedure was continued with balloon angioplasty.A delay of 30 minutes or greater occurred due to this issue.
|
| |
|
Search Alerts/Recalls
|