| Catalog Number |
MC1352P |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2020-00173.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, an roi-c fractured during plate installation; the plate would not advance without several very hard hits with a mallet.The cage and plate were removed from the wound and replaced with an alternative cage and plate to complete the procedure.It was also reported that x-rays were not taken prior to plate installation to ensure proper cage placement and distance from the retraction pins.There were no reported patient impacts associated with this event.This is report one of two for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, an roi-c fractured during plate installation; the pate would not advance without several very hard hits with a mallet.The cage and plate were removed from the wound and replaced with an alternative cage and plate to complete the procedure.It was also reported that x-rays were not taken prior to plate installation to ensure proper cage placement and distance from the retraction pins.There were no reported patient impacts associated with this event.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information and initially corrected information.Information added to d10, h3, and h6: method, results, and conclusions.The complaint is confirmed for one (1) of one (1) roi-c cage (pn: mc1352p) for the failure of cracked.Medical records were not provided for review.Device evaluation: visual inspection of the cage shows that it has fractured.The complaint is confirmed.Potential cause root cause was unable to be determined.Dhr review and related actions per dhr review, the parts were likely conforming when they left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
| |
|