Model Number DXT1612AL
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic laparoscopic transabdominal preperitoneal, the mesh was damaged when the surgeon attempted to insert it through the robotic 8mm trocar.The mesh was not implanted.The surgeon used a larger diameter trocar and another mesh was successfully implanted and the case was completed.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the sample was not returned in its original packaging but placed in a bubble wrap envelop.The sample was found not contaminated by blood and was received folded in 4 and maintains this fold.The textile knitting pattern and the mesh dimension were found as expected.At the bend points, the sealed edges were damaged and cracked.The textile was damaged on 3x3 pores, above the ml green marking.Also, the mesh was placed with robotic assistance and used a trocar of 8 mm.It was reported that the mesh tore.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: it is recommended to use a trocar of at least 10 mm internal diameter to introduce meshes of size 13x09 cm and 15x10 cm, and a trocar of at least 11 mm internal diameter to introduce a mesh of size 16x12 cm.Mesh insertion capability may vary depending on mesh insertion method, mesh size and graspers/trocars used.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXTILE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
MDR Report Key10760048
Report Number9615742-2020-02438
Device Sequence Number19672705
Product Code FTL
UDI-Device Identifier10884521726543
UDI-Public10884521726543
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K192443
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberDXT1612AL
Device Catalogue NumberDXT1612AL
Device Lot NumberSUF1189X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Initial Date Received by Manufacturer 10/08/2020
Supplement Date Received by Manufacturer12/08/2020
Initial Report FDA Received Date10/30/2020
Supplement Report FDA Received Date12/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-