|
According to the reporter, during a robotic laparoscopic transabdominal preperitoneal, the mesh was damaged when the surgeon attempted to insert it through the robotic 8mm trocar.The mesh was not implanted.The surgeon used a larger diameter trocar and another mesh was successfully implanted and the case was completed.There was no patient injury.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the sample was not returned in its original packaging but placed in a bubble wrap envelop.The sample was found not contaminated by blood and was received folded in 4 and maintains this fold.The textile knitting pattern and the mesh dimension were found as expected.At the bend points, the sealed edges were damaged and cracked.The textile was damaged on 3x3 pores, above the ml green marking.Also, the mesh was placed with robotic assistance and used a trocar of 8 mm.It was reported that the mesh tore.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: it is recommended to use a trocar of at least 10 mm internal diameter to introduce meshes of size 13x09 cm and 15x10 cm, and a trocar of at least 11 mm internal diameter to introduce a mesh of size 16x12 cm.Mesh insertion capability may vary depending on mesh insertion method, mesh size and graspers/trocars used.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|