Model Number NAV2132
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with preoperative diagnosis for spondylolithesis therapy.It was reported that the inserter was broken.There was no fragment left or contact with the patient.There were no symptoms reported.There were no further complications reported regarding the event.The inserter shaft broke from the inserter handle when the physician unscrewing the inner inserter tube to release the inserter from the interbody.
 
Additional Manufacturer Narrative
H3: product id:nav2132 lot# dn17a001 visual and optical examination identified that the elevate inserter has broken on the outer shaft where the shaft and handle interface.The break has occurred at the locking tab cutouts.A microscopic analysis of the fracture surface indicates a brittle fracture.The fracture appears to have occurred from the material attempting to shift during compression.This is consistent with the instrument being struck from the back side.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE SPINAL SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10760515
Report Number1030489-2020-01508
Device Sequence Number3185068
Product Code OLO
UDI-Device Identifier00643169771000
UDI-Public00643169771000
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K163581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNAV2132
Device Catalogue NumberNAV2132
Device Lot NumberDN17A001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2020
Supplement Date Received by Manufacturer01/13/2021
Initial Report FDA Received Date10/30/2020
Supplement Report FDA Received Date01/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
Patient Weight122
-
-