• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Pain (1685); Dehydration (1807); Dyspnea (1816); Hyperglycemia (1905); Nausea (1970); Vomiting (2144); Diabetic Ketoacidosis (2364)
Event Date 10/22/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer's wife reported via phone call that they were hospitalized for high blood glucose, diabetic keto acidosis on (b)(6) 2020 with blood glucose level was 1074 mg/dl.Customer experienced symptoms such as nausea, vomiting, abdominal pain, difficulty in breathing, dehydration.Customer was treated with insulin drip.Customer did not feel ok to troubleshooting for high blood glucose level.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test.Device received with pillowing keypad overlay.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10761196
MDR Text Key213846370
Report Number2032227-2020-198906
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000166519
UDI-Public(01)000000763000166519(17)220922
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/22/2022
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG3RWGF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Date Manufacturer Received11/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-MMT-332,UNOMED INF SET
Patient Outcome(s) Hospitalization;
Patient Age53 YR
Patient Weight274
-
-