The customer's wife reported via phone call that they were hospitalized for high blood glucose, diabetic keto acidosis on (b)(6) 2020 with blood glucose level was 1074 mg/dl.Customer experienced symptoms such as nausea, vomiting, abdominal pain, difficulty in breathing, dehydration.Customer was treated with insulin drip.Customer did not feel ok to troubleshooting for high blood glucose level.The insulin pump will be returned for analysis.
|
Device passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test.Device received with pillowing keypad overlay.(b)(4).
|