Model Number 5510-F-501
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Device evaluated by manufacturer? not returned.
 
Event or Problem Description
Surgeon selected needed implants after trialing.I selected implants off the knee implant cart.There was a cemented femur implant on the cart where the corresponding cementless implant was designated.All implants were shown to the circulator and surgeon prior to being placed on the field.Cemented femur was implanted, rather than the intended cementless implant.The patient was brought back into the room, and the cemented implant was removed and replaced with the proper cementless implant.
 
Additional Manufacturer Narrative
Reported event an event regarding incorrect selection involving a triathlon femoral component was reported.The event was confirmed via review of the provided implant sheet.Method & results -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-medical records received and evaluation: no medical records were received for review with a clinical consultant.-product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported a cemented implant was implanted, rather than the intended cementless implant.Provided implant sheet was reviewed and confirmed that the reported device (catalog # 5510-f-501) was implanted without bone cement.The reported event was confirmed.It was also reported that the incorrect implant was shown to the circulator and surgeon prior to being placed on the field.It was an user error that the incorrect implant was used.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.Device not returned.
 
Event or Problem Description
Surgeon selected needed implants after trialing.I selected implants off the knee implant cart.There was a cemented femur implant on the cart where the corresponding cementless implant was designated.All implants were shown to the circulator and surgeon prior to being placed on the field.Cemented femur was implanted, rather than the intended cementless implant.The patient was brought back into the room, and the cemented implant was removed and replaced with the proper cementless implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #5 L-CEM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10761646
Report Number0002249697-2020-02283
Device Sequence Number11034372
Product Code MBH
UDI-Device Identifier07613327040296
UDI-Public07613327040296
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number5510-F-501
Device Catalogue Number5510F501
Device Lot NumberBE97E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/07/2020
Supplement Date Received by Manufacturer11/23/2020
Initial Report FDA Received Date10/30/2020
Supplement Report FDA Received Date12/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age52 YR
-
-