|
It was reported that the delivery wire got broken.The target lesion area was located in a moderately tortuous right femoral artery.A 14mmx50cm interlock embolic was selected for use in a coil embolization.During the procedure, it was noted that the device got stuck inside the microcatheter and could not advance.However, the delivery wire got broken/kinked.The device was removed together with the microcatheter.The procedure was completed with another of the same device.There were no patient complications reported.
|