| Catalog Number |
14051454200 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
09/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation revealed that this incident was user related.The staff did not follow the safety instructions detailed in the instructions for use (ifu).The customer did not wear cut-resistant gloves while cleaning the instrument, as described in the ifu.A customer facing letter will be sent out with recommendation in the future to follow the cleaning and maintenance instructions provided in the instructions for use and to wear cut-resistant gloves.
|
| |
|
Event or Problem Description
|
|
On 02 october 2020, leica biosystems received a complaint that a user was injured while attempting to clean the shifter and rail on their coverslipper, histocore spectra cv.The injured user had pieces of glass in her finger and visited the emergency room to have the glass removed and then returned to work.No stitches were required.The user was identified as a female in her 20's.
|
| |
|