• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL PALMTOP VENTILATOR, MODELS PTV-8, PTV-10; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL PALMTOP VENTILATOR, MODELS PTV-8, PTV-10; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 19260-001
Device Problem No Pressure (2994)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Results of investigation: the vyaire factory service dept.Received the revel ventilator and performed an investigation.The ventilator was powered on and no flow was being produced and no breaths were being delivered.Bench testing revealed the blower module was malfunctioning.The reported problem was confirmed.Inspection of the blower module revealed the component q4 was burned.The device was repaired, passed final testing and was returned to the customer.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported to vyaire medical that the revel ventilator was not producing any pressure or air during pretesting.The customer advised there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALMTOP VENTILATOR, MODELS PTV-8, PTV-10
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
17400 medina road suite 100
minneapolis MN 55447
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key10763074
MDR Text Key213902937
Report Number2031702-2020-03720
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00845873000029
UDI-Public(01)00845873000029(11)20180108
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070594
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number19260-001
Device Catalogue Number19260-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-