| Model Number |
9770602X |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
| |
|
Event or Problem Description
|
|
Per biomed, they believe there is a cracked crystal due to the lines in the image and poor image quality.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of lines in the image and poor image quality was confirmed; there are black vertical lines showing on the display.The root cause of the reported failure was identified as a probe failure.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|
| |
|
Event or Problem Description
|
|
Per biomed, they believe there is a cracked crystal due to the lines in the image and poor image quality.
|
| |
|