| Model Number |
SZ282R |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
According to the complaint description the surgeon used a hammer to seat the counter torque and there was no impact to patient.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2020-00759 (b)(4).9610612-2020-00760 (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Associated medwatch-reports: 9610612-2020-00759 ((b)(6) - sz282r), 9610612-2020-00760 ((b)(6) - sz277r).Due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.In similar cases as the described, the rod persuader was not completely / correct attached at the pedicle screw.This leads to the jamming of the instrument.The ifu figures out, to ensure the correct attaching to the head of the pedicle screw.Therefore we assume a handling error as the most likely root cause.
|
| |
|