MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 3-4 LT 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 3-4 LT 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74023841
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Information (3190)
Date of Event 10/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tka performed on (b)(6) 2018, the patient experienced an unspecified adverse event that was addressed by performing a revision surgery: all components were explanted.The patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms per complaint details, a 1st tka revision was performed approximately 2.5 years post implantation.It was communicated that the requested medical documentation would not be provided; therefore, the root cause of the reported event could not be fully assessed.The patient impact beyond the reported revision could not be determined.No further medical assessment is warranted at this time.Should additional information/ documentation become available, the clinical/medical task may be re-opened for further evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II XR INS XLPE LAT 3-4 LT 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10766945
Report Number1020279-2020-05976
Device Sequence Number14249302
Product Code JWH
UDI-Device Identifier00885556615669
UDI-Public00885556615669
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74023841
Device Catalogue Number74023841
Device Lot Number17JM18494
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/05/2020
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date10/30/2020
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-