MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 STANDARD 9MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 STANDARD 9MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035469300
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Cerebral Edema (4403); Thrombosis/Thrombus (4440)
Date of Event 10/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device not returned for evaluation.
 
Event or Problem Description
It was reported that during the procedure to treat subarachnoid hemorrhage, the coil (subject device) got stretched and detached prematurely at the aorta artery.The physician attempted to remove the coil with a snare device, but was unsuccessful.It was reported that the patient passed away after the procedure due to brain swelling caused by thrombogenic progression and spasm developed due to the preexisting subarachnoid hemorrhage.
 
Event or Problem Description
It was reported that during the procedure to treat subarachnoid hemorrhage, the coil (subject device) got stretched and detached prematurely at the aorta artery.The physician attempted to remove the coil with a snare device, but was unsuccessful.It was reported that the patient passed away after the procedure due to brain swelling caused by thrombogenic progression and spasm developed due to the preexisting subarachnoid hemorrhage.
 
Additional Manufacturer Narrative
H4 manufacturing date: added.D4 expiration date: added.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that the patient had subarachnoid hemorrhage (sah) before starting the procedure, and the blood vessels had a lot of thrombogenic progression and spasm.The patient's blood vessels were very long and tortuous.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to the as reported.H3 other text : device not returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 STANDARD 9MM X 30CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
MDR Report Key10767041
Report Number3008881809-2020-00321
Device Sequence Number3645673
Product Code HCG
UDI-Device Identifier04546540676351
UDI-Public04546540676351
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2022
Device Model NumberM0035469300
Device Catalogue NumberM0035469300
Device Lot Number22061691
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer11/30/2020
Initial Report FDA Received Date10/30/2020
Supplement Report FDA Received Date12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 CATHETER (STRYKER).
Outcome Attributed to Adverse Event Death; Required Intervention;
-
-