Model Number 72200873
Device Problems Overheating of Device (1437); Electrical Shorting (2926); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2020
Event Type  malfunction  
Event Description
It was reported that during surgery, the dii controller was not functioning, when the motor drive unit was plug in to port a it gave the error message "short circuit detected", the same issue happened when plugging into port b.The hand piece was defective due to an intermittent open circuit/or break in the wiring.The controller was extremely warm and a burning smell was noticed.The procedure was completed without delay using a s+n backup device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of "short circuit" error message was confirmed.Product failed functional testing with a short circuit error in port a only.Cause of errors is a defective electronic component on the main digital pcb.Product passed functional testing with a known good main pcb installed.It appears that a shorted hand piece was inserted into unit.The complaint was confirmed and the root cause has been determined to be a defective electronic component.A review of manufacturing records found the device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10767264
MDR Text Key214104610
Report Number1643264-2020-01664
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010607409
UDI-Public03596010607409
Combination Product (y/n)N
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200873
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2020
Date Manufacturer Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-