MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY PDX, INTEGRATED SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY PDX, INTEGRATED SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number 050-87222
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the cassette door of their cycler during step 9 of their pd end treatment.The patient reported receiving an air detected in cassette alarm during fill 2 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The fluid leak did come in contact with the cycler.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was not available.Additional information was requested, however; to date has not been provided.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the cassette door of their cycler during step 9 of their pd end treatment.The patient reported receiving an air detected in cassette alarm during fill 2 of 5 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The fluid leak did come in contact with the cycler.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was not available.Upon follow up, the pdrn reported the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The pdrn stated that treatment was completed after receiving the new cycler.The pdrn reported that it is unknown where exactly the fluid leak occurred from.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient has received a new cycler which is working well and is continuing peritoneal dialysis therapy with no further issues.The liberty cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.The cycler was returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY PDX, INTEGRATED SET
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
MDR Report Key10767273
Report Number8030665-2020-01607
Device Sequence Number11574628
Product Code FKX
UDI-Device Identifier00840861101993
UDI-Public00840861101993
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K141145
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2021
Device Model Number050-87222
Device Catalogue Number050-87222
Device Lot Number20AR08128
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Received by Manufacturer 10/14/2020
Supplement Date Received by Manufacturer11/17/2020
Initial Report FDA Received Date10/31/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; LIBERTY PDX CYCLER
Patient Age83 YR
Patient Weight91
-
-