Model Number 9445-02 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that signal loss over one hour occurred.A review of the share logs was performed and signal loss was found within the investigation window.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
|
Manufacturer Narrative
|
(b)(4).B5: describe event or problem additional.D10: device available for evaluation additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer additional.H6: event problem and evaluation codes additional.
|
|
Event Description
|
It was reported that signal loss over one hour occurred.The product was evaluated.An external/exterior visual inspection was performed and passed.Voltage check was performed and passed.Pairing test/download was performed and passed.A review of the share log was performed and signal loss was found within the investigation window.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
|