• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201186
Medical Device Problem Codes Difficult to Open or Close (2921); Mechanical Jam (2983)
Health Effect - Clinical Codes No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during arthroscopy, the arthropierce instrument's grabbing arm stopped closing in mid procedure.There was a delay of under 30 min, the procedure was completed with a change in technique using a competitor device (mitek cleverhook).No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review for the reported batch number concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROPIERCE INSTRUMENT CURVED LEFT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10767698
Report Number1219602-2020-01720
Device Sequence Number19714102
Product Code NBH
UDI-Device Identifier03596010605610
UDI-Public03596010605610
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201186
Device Catalogue Number72201186
Device Lot Number50599989
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2020
Supplement Date Received by Manufacturer05/21/2021
Initial Report FDA Received Date11/01/2020
Supplement Report FDA Received Date06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-