| Model Number |
7207561 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a procedure, the biorci anchor broke while inserting it into tibia.There was a backup device available.No delay or further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H10, h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found: the starter must be utilized with the biorci screws to minimize screw breakage during insertion.Excessive force should not be placed on the delivery instrument.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|
Search Alerts/Recalls
|
|
|