Model Number 2342305M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider, user facility) regarding a patient had previous fusion (tlif).It was reported that the bone was hard and the driver sheared off when doctor tried to upsize the screw.The product will be returned.There were no patient symptoms or complications as a result of this event.
 
Additional Manufacturer Narrative
H6: product analysis: visual and optical examination confirmed approximately 3mm of instrument tip has been broken off, consistent with interface during usage.Optical examination revealed a fairly flat surface with circular material flow.The remaining tip has been twisted.This type of damage is consistent with torsional overload.Additional information: d9, g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWEREASE INSTRUMENTS
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10768376
Report Number1030489-2020-01529
Device Sequence Number16422491
Product Code HBE
UDI-Device Identifier00613994905697
UDI-Public00613994905697
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K111520
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2342305M
Device Catalogue Number2342305M
Device Lot NumberCT16L039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2020
Supplement Date Received by Manufacturer02/04/2021
Initial Report FDA Received Date11/02/2020
Supplement Report FDA Received Date02/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-