| Model Number |
2990003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Radiographic image review: provided radiographic image analysis found that at the infusion level of the construct, there appears to be a metallic fragment adjacent to the interbody device, this is reported as a tip of inserter which likely fractured during attempted torsional movement of the shaft.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare professional via manufacturer representative regarding a patient (medical history-diabetes) having plif (posterior lumbar interbody fusion) (levels implanted -l3/l5) due to lumbar canal stenosis (lcs).It was reported that after the cage was inserted, the tip of the inserter broke due to a slight twist to adjust the position.The grip part at the tip was broken.The tip of the inserter remained on the grip part of the cage.Although the cage was attempted to be removed, it was difficult, so the cage was left as it was.There were no patient symptoms as a result of the event.The product return was requested.There was a procedural delay of less than 60 mins.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis:2990003, lot# nm12a027 visual and optical inspection of the threaded inserter did not reveal any non-conformances.Functionally tested the threaded inserter with a sample 2990001 inserter with no issues.No fault found.Inserter is capable of performing its intended function.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|