BIOMET 3I DO NOT USE - CERTAIN PREVAIL IMPLANT 5/4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number IIOS5411
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Unspecified Infection (1930); Inflammation (1932); No Code Available (3191)
Date of Event 09/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).Patient identifier unknown / not provided.Weight unknown / not provided.Reporter name and email were not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site 19 was removed due to a peri-implantitis.Symptoms as a result of the event: abcess & inflammation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DO NOT USE - CERTAIN PREVAIL IMPLANT 5/4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10770029
Report Number0001038806-2020-01715
Device Sequence Number1901742
Product Code DZE
UDI-Device Identifier00844868005902
UDI-Public(01)00844868005902(17)190321(10)2014030467(241)IIOS5411
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K061629
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/21/2019
Device Model NumberIIOS5411
Device Catalogue NumberIIOS5411
Device Lot Number2014030467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/08/2020
Supplement Date Received by Manufacturer02/05/2021
Initial Report FDA Received Date11/02/2020
Supplement Report FDA Received Date02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-