| Model Number |
NSLG2C25 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # r94l5j.Investigation summary: the device was received with the top of the i-blade upper tip broken off not returned.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total abdominal hysterectomy the enseal jaws did not close and did not open.The procedure was completed with no patient consequences.
|
| |
|