| Model Number |
NONB5SHF |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, when the trocar was introduced, plastic particles were detached from it, and adhered to the wall (peritoneum) of the patient.The same trocar having been used for the placement of the other cannulas (3 shorts of 5mm in total), residues plastics are also observed at the introduction sites.It was also reported that persistence of plastic in the patient with inability to remove plastic debris.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but photos were available for evaluation.The first image showed the distal end of the obturator.There were visible gouge marks on the obturator.The second image appeared to be an x-ray and there were light pink and blue hues.It was reported that a component disengaged from the device into the surgical cavity and the obturator was damaged.The reported issues were confirmed.The most likely cause could not be established from the information available.Gouges in the obturator can occur if device was handled roughly during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic procedure, when the trocar was introduced, plastic particles were detached from it and adhered to the wall (peritoneum) of the patient.The same trocar having been used for the placement of the other cannulas (3 shorts of 5mm in total), residues plastics are also observed at the introduction sites.It was also reported that persistence of plastic in the patient with inability to remove plastic debris.
|
| |
|