• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC VERTECEM V+ SYRINGE KIT; CEMENT DISPENSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC VERTECEM V+ SYRINGE KIT; CEMENT DISPENSER Back to Search Results
Model Number 03.702.215S
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2020 during a vertebroplasty procedure, when the surgical cement was loaded into the 1cc syringes it propelled from the plunger.This prevented loading the cement for injection.There was a surgical delay of five (5) minutes.The procedure was successfully completed with the use of alternative 2cc syringes.There was no patient consequence.Concomitant device reported: unknown cement (part#: unknown, lot#: unknown, quantity: 1) this report is for one (1) vertecem v+ syringe kit.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part 03.702.215s, lot 8022529: manufacturing site: selzach.Supplier: (b)(4).Release to warehouse date: march 13, 2018.Expiry date: february 01, 2023.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: a product investigation was completed: upon visual inspection, it is observed that one blue connector and thirteen 1ml syringes were received (5 white, 8 blue).One of the blue syringe plungers was stuck due to hardened cement.One of the blue syringe plungers was disconnected from body.No defect was observed with the rest of the syringes received.Functional test was be performed for the syringes with the hardened cement.Rest of the syringes received were working normally as intended.But the reported leak condition cannot be confirmed.The relevant drawings were reviewed; no design issues or discrepancies were noticed.The complaint condition was not confirmed.There were no issues during the manufacture of this product that would contribute to this complaint condition.A definitive root cause cannot be determined for the reported problem.During the investigation, no product design issues, or discrepancies were observed that may have contributed to the complaint condition.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTECEM V+ SYRINGE KIT
Common Device Name
CEMENT DISPENSER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10771382
Report Number2939274-2020-04926
Device Sequence Number2113383
Product Code KIH
UDI-Device Identifier10705034721699
UDI-Public(01)10705034721699
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2023
Device Model Number03.702.215S
Device Catalogue Number03.702.215S
Device Lot Number8022529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer11/02/2020
12/08/2020
Initial Report FDA Received Date11/02/2020
Supplement Report FDA Received Date11/25/2020
12/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - CEMENT
-
-