| Model Number |
6L45-41 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer obtained a falsely elevated architect total bilirubin result for one patient.Sample id (b)(6) generated 8.49 and repeat 0.21 mg/dl.A serum sample from the same patient was also tested, and generated 0.18 mg/dl.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the customers issue included a review complaint data, trending, lot search and labeling.A review of tickets determined that there is normal complaint activity for lot 56892uq03.Review of trending reports did not identify any issues.Review of manufacturing documentation for the complaint lot found no issues related to the current complaint.Review of product labeling found adequate information to address the complaint issue.The information provided by the customer indicated the same patient sample and a new sample were retested with similar expected lower results.In addition, the qc results were within expected ranges indicating the product is performing as expected.Based on all available information, no systemic issue or deficiency of the architect total bilirubin assay was identified.
|
| |
|