| Model Number |
8577 |
| Medical Device Problem Codes |
Device Contamination with Chemical or Other Material (2944); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned to boston scientific for analysis.Returned product consisted of a mamba flex microcatheter.The hub, shaft and tip were microscopically examined.The device showed a damaged area approximately 65.5cm from the tip.The foreign material the customer returned was confirmed.The returned rota wire was inserted into the lumen of the device.The wire transcended through the device until it met resistance approximately 115cm from the hub.The device was x-rayed; however, nothing appeared on the x-ray image.The wire was removed, and the device was soaked for a period of 24 hours until the lumen blockage was cleared.A syringe of fluid was injected through the device to show patency.It was noticed that the damaged area inflated like a balloon which gives the indication of shaft damage and ptfe inner liner damage.The wire was reinserted and transcended through the device with a slight sticking point at the 65.5cm damage.The foreign material that was recovered was tested on the ftir testing equipment.The foreign material that was noticed is consistent with the wire interfering with the ptfe liner.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.No other issues were identified during the product analysis.The complaint was confirmed for difficulty loading the wire and foreign material.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was present in the device.A percutaneous coronary intervention was being performed to treat a blockage in the ostial right coronary artery.A rota wire was being used and would not advance through the mamba flex microcatheter.The catheter was removed and flushed and a hair-like substance came out.The mamba device was not reinserted in the body.The rota wire was used to complete the procedure.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|