Model Number 3CX*FX25REC
Device Problems Crack (1135); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2020
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation, however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the arterial outlet is deformed and cracked.No patient involvement; product was changed out; procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 2, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Method code #1: 10 testing of actual/suspected device.Method code #2: 11 testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 analysis of production records.Results code: 3259 improper physical structure.Conclusions code: 4307 cause traced to component failure.The returned complaint sample was investigated and found to have damage to the port.The port did have a crack and was dented at the opening of the port.A representative retention sample from the same lot number was inspected and found to be free of damage.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key10773195
MDR Text Key214798631
Report Number1124841-2020-00253
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberYF25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Was the Report Sent to FDA? No
Date Manufacturer Received12/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-