It was reported that, five weeks after a hemiarthroplasty had been performed, the patient experienced pain after a dislocation that occurred because of prohibited motion.The patient suffers from cognitive impairment.A revision surgery was performed to treat the adverse event: although the surgeon confirms that no s+n devices were defective, the tndm bp shl/xlpe liner 54od 26id component was exchanged.The current status of patient is not known.
|
G3, h2,h3 and h6: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms this case reports a revision was performed due to a dislocation that occurred as a result of prohibited motion.Per email communication, the requested surgical reports and x-rays are not available.It was also communicated that the surgeon confirmed that the devices were not defective.Therefore, patient impact beyond the revision cannot be determined.No further medical assessment is warranted at this time.Should clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
|