• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 1012535-100
Medical Device Problem Codes Difficult or Delayed Activation (2577); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual and functional analysis was performed on the returned device.The reported difficulties were unable to be confirmed as the stent was already fully deployed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar complaints.The investigation was unable to determine cause for the reported difficulties.It may be possible that the distal shaft was bent or restricted in the anatomy preventing the shaft lumens from moving freely and causing the resistance with the thumbwheel; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was a reported that the procedure was to treat a lesion located in the proximal superficial femoral artery.Access to the vessel was obtained via antegrade approach, right superficial artery using 6f 45 cm sheath.The handle was unlocked.The physician went to rotate the thumbwheel with great resistance, and the stent began to un-flower so he kept rotating to fully deploy the stent.The physician stated there was lots of resistance with the thumbwheel; however, the stent was successfully deployed in the correct location.There were no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10774990
Report Number2024168-2020-09139
Device Sequence Number16959364
Product Code NIP
UDI-Device Identifier08717648176098
UDI-Public08717648176098
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model Number1012535-100
Device Catalogue Number1012535-100
Device Lot Number0072263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/14/2020
Initial Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight119
-
-