| Model Number |
BXA112902A |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium apl, which in covalently bound to the device surface and is essentially|non-eluting.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a gore® viabahn® vbx balloon expandable endoprosthesis(vbx) was being placed as a innominate stent in conjunction with transcarotid artery re-vascularization (tcar) procedure.During the advancement of the device the physician had to use excessive force to get the device into place.After the device was successfully deployed, the physician was removing the device catheter through the 8fr tcar introducer sheath, and felt resistance.After the resistance was felt he continued to withdraw the catheter, and the catheter broke.The physician removed the remainder of the catheter and sheath in tandem.When it was removed, it was suspected that the balloon portion of the catheter hung up on the sheath tip.Patient did not suffer any adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: component code 515 (stent) added.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.A product history review was conducted for the vbx device.The manufacturing records were reviewed, and the device lot met all pre-release specifications.Assessment of any potential failure modes and causes is not possible because the endo remains implanted and the delivery catheter was not returned.Cause for the reported insertion and withdrawal difficulty and the broken delivery catheter cannot be established.
|
| |
|
Search Alerts/Recalls
|