| Model Number |
9697 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that her blood glucose readings were not being stored in the memory of the contour next meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next meter and contour next test strips for evaluation.The returned meter was tested with the returned test strips using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly stored in the meter's memory.Section d4 of the initial report indicated the serial # of the suspected contour next meter as (b)(6).Based on the updated information, the correct serial # for the suspected meter is (b)(6).Section d4 has been updated with the correct serial #.The device manufacture date (h4) remains the same.
|
| |
|