Model Number 71175068
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during the procedure, that the tip of the depth gauge was broken outside the patient.The procedure was completed without delay with a backup from smith & nephew.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The device, intended for use in treatment was returned for evaluation: a visual inspection of the returned device confirmed the stated failure mode.The tip of the device is heavily damaged.A review of complaint history did reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 3.5MM DEPTH GAUGE LONG
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10778167
Report Number1020279-2020-06064
Device Sequence Number1126427
Product Code HRS
UDI-Device Identifier00885556635865
UDI-Public00885556635865
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K162078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175068
Device Catalogue Number71175068
Device Lot Number18KRS0028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date11/03/2020
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-