| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
Device returned with packaging damage and sterility barrier compromised.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample wa evaluated and the reported event was confirmed.Visual inspection of the returned products found to exhibit damage to the cartons, outer blisters are cracked, and inner blisters are exhibit stress marks.Sterility has been compromised.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the product when it left zimmer biomet control was conforming to specifications.The root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|